What are the sterilization guidelines for the new electric handpieces? Does the motor have to be sterilized?

Ask OSAP does not review, evaluate, certify, recommend or endorse products. Ask OSAP also does not provide technical support for specific products. If you have further questions about procedures and specific products, it is recommended that you consult with the manufacturer’s written instruction manual and/or contact the manufacturer of the electric handpiece directly. The manufacturer can best advise you regarding the electric handpiece’s instructions for use and maintenance (IFU). Failure to comply with the manufacturer’s instructions for use could potentially void the product warranty.

In general, the 2003 CDC guidelines for infection control in dentistry states as follows:

Heat methods can sterilize dental handpieces and other intraoral devices attached to air or waterlines (246,275,356, 357,360). For processing any dental device that can be removed from the dental unit air or waterlines, neither surface disinfection nor immersion in chemical germicides is an acceptable method. Ethylene oxide gas cannot adequately sterilize internal components of handpieces (250,275). In clinical evaluations of high-speed handpieces, cleaning and lubrication were the most critical factors in determining performance and durability (361–363). Manufacturer’s instructions for cleaning, lubrication, and sterilization should be followed closely to ensure both the effectiveness of the process and the longevity of handpieces. 1

Resources

1)  Kohn WG, Collins AS, Cleveland JL, Harte JA, Eklund KJ, Malvitz DM, Centers for Disease Control and Prevention (CDC). Guidelines for infection control in dental health-care settings—2003. MMWR Recomm Rep 2003;52(RR-17):1-61. http://www.cdc.gov/mmwr/preview/mmwrhtml/rr5217a1.htm    Accessed on October 26, 2015.